Comparison of scansys and pentacam HR in healthy eyes

Share:

Hesam Hashemiana, Mehdi Khodaparasta, Masoud Khorrami-Nejadb,c, Marrwan Hisham 

Mohmmedb,c, Hooman Ahmadzadeha, Majid Akhzari Hamedania, Yasir Adil Shakora and Elham Azizid

a Translational Ophthalmology Research Center, Farabi Eye Hospital, Tehran, Iran;

b Optometry Department, School of Rehabilitation, Tehran University of Medical Sciences, Tehran, Iran;

c Department of Optical Techniques, Al Mustaqbal University College, Hillah, Babylon, Iraq;

d Department of Optometry and Vision Science, University of Melbourne, Parkville, Victoria, Australia


Background: To determine the interchangeability of Scansys (MediWorks, Shanghai, China) and pentacam HR (Oculus, Wetzlar, Germany) in anterior segment measurements.

Research designs and methods: 42Forty-two healthy individuals’ right eye (mean age 31.5 ± 7.3 years) were assessed for keratometry (K), central corneal thickness (CCT), thinnest corneal thickness (TCT), pupil center pachymetry, corneal volume, anterior chamber volume (ACV), anterior chamber depth (ACD), terior chamber angle (ACA), intraocular pressure (IOP) sum, Q-value (asphericity), and horizontal white-to-white (WTW). Intraclass correlation coefficient (ICC), linear regression analysis, difference of mean, and Bland-Altman plots [bias line and 95% limits of agreement (LoA)] were analyzed.

Results: Scansys repeated excellently the Pentacam HR for keratometry (except for anterior steep K), ACD, pupil center pachymetry, CCT, TCT, IOP sum, and corneal volume (all ICC > 0.9, r 2 > 0.8, and narrow LoAs). There was strong correlations in measuring WTW and ACV (r2 = 0.631 and r 2 = 0.623, respectively) and weak correlations in measuring ACA and Q values (all r 2 < 0.5).

Conclusion: The interchangeability of devices were confirmed in measuring ACD, pupil center achymetry, CCT and TCT, IOP sum, corneal volume, and keratometry except for the anterior steep K. However, in measuring WTW, ACV, asphericity, and ACA, Scansys should not be used instead of Pentacam HR.